Pular para o conteúdo principal
Alarm — stator control board error conditionSlip Ring Communication

ER20Slip ring send/receive signal short circuit

Philips CT error code, reported by the Slip Ring Communication.

What ER20 means

On Access Series CT (6/16/Dual) CT systems, status code ER20 indicates: Slip ring send/receive signal short circuit. The code originates in the Slip Ring Communication section of the Slip Ring Communication. Severity level alarm (Alarm — stator control board error condition).

Reported action: Cross-reference System Error Code 0x5711 in the system log. Inspect the slip ring transmitter/receiver signal wiring for a short before replacing the affected module..

Listed in Table 1-3 'Stator Control Board Error Code vs System Error Code' of the Philips Access CT Service Trouble Shooting Guide (459800863401 Rev. C), for the Access Series CT (6/16/Dual) family. Not confirmed for other Philips CT systems.

Full troubleshooting for ER20

The complete diagnosis, cause tree, and step-by-step remedy for the Slip Ring Communication subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related Slip Ring Communication codes

Discuss ER20 with other field engineers in the FSELIB forum

Source: Philips Access CT Service Trouble Shooting Guide — 459800863401 Rev. C. Reference information re-authored by FSELIB — not a copy of the OEM manual.