Alarm — stator control board error conditionPatient Table / Patient Support

ER1APatient table CAN communication time-out

Philips CT error code, reported by the Patient Table / Patient Support.

What ER1A means

On Access Series CT (6/16/Dual) CT systems, status code ER1A indicates: Patient table CAN communication time-out. The code originates in the Patient Table / Patient Support section of the Patient Table / Patient Support. Severity level alarm (Alarm — stator control board error condition).

Reported action: Cross-reference System Error Code 0x5705 in the system log. Check the CAN cable/connector between the Patient Table Motion Control Board (CAN ID 3, Table 1-6) and the Stator Control Board before replacing either board or the cable..

Listed in Table 1-3 'Stator Control Board Error Code vs System Error Code' of the Philips Access CT Service Trouble Shooting Guide (459800863401 Rev. C), for the Access Series CT (6/16/Dual) family. Not confirmed for other Philips CT systems.

Full troubleshooting for ER1A

The complete diagnosis, cause tree, and step-by-step remedy for the Patient Table / Patient Support subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related Patient Table / Patient Support codes

Discuss ER1A with other field engineers in the FSELIB forum

Source: Philips Access CT Service Trouble Shooting Guide — 459800863401 Rev. C. Reference information re-authored by FSELIB — not a copy of the OEM manual.