Alarm — critical, operation haltedProcessor (CPU)

E1414Microprocessor exceeded the acceptable temperature range and halted operation (Thermtrip)

Philips CT error code, reported by the MND Extended Storage Server (Dell PowerEdge 2900 hardware).

What E1414 means

On Mobile Network Device (MND) Extended Storage Option CT systems, status code E1414 indicates: Microprocessor exceeded the acceptable temperature range and halted operation (Thermtrip). The code originates in the Processor (CPU) section of the MND Extended Storage Server (Dell PowerEdge 2900 hardware). Severity level alarm-high (Alarm — critical, operation halted).

Reported action: If the problem persists, ensure the microprocessor heat sinks are properly installed (see Troubleshooting the Microprocessors, p.82). Per the guide, the LCD keeps showing this message until the power cord is disconnected/reconnected or the SEL is cleared via Server Assistant or the BMC Management Utility..

Listed as an LCD front-panel status/error code ('CPU # Thermtrip', Table 2 LCD Status Messages, p.21). '#' identifies the affected processor.

Full troubleshooting for E1414

The complete diagnosis, cause tree, and step-by-step remedy for the Processor (CPU) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.