Alarm — servo drive faultPatient Table / Patient Support

ALE20Patient table servo drive serial connection time-out

Philips CT error code, reported by the Patient Table Servo Drive.

What ALE20 means

On Access Series CT (6/16/Dual) CT systems, status code ALE20 indicates: Patient table servo drive serial connection time-out. The code originates in the Patient Table / Patient Support section of the Patient Table Servo Drive. Severity level alarm (Alarm — servo drive fault).

Reported action: Check the serial communication cable/connector between the servo drive and the Patient Table Motion Control Board for an intermittent or slow link before replacing either unit..

Listed in Table 1-12 'Patient Table Servo Error Codes' of the Philips Access CT Service Trouble Shooting Guide (459800863401 Rev. C), for the Access Series CT (6/16/Dual) family. Not confirmed for other Philips CT systems.