ALE19 — Patient table servo drive serial connection abnormal
Philips CT error code, reported by the Patient Table Servo Drive.
What ALE19 means
On Access Series CT (6/16/Dual) CT systems, status code ALE19 indicates: Patient table servo drive serial connection abnormal. The code originates in the Patient Table / Patient Support section of the Patient Table Servo Drive. Severity level alarm (Alarm — servo drive fault).
Reported action: Check the serial communication cable/connector between the servo drive and the Patient Table Motion Control Board for damage or a loose connection before replacing either unit..
Listed in Table 1-12 'Patient Table Servo Error Codes' of the Philips Access CT Service Trouble Shooting Guide (459800863401 Rev. C), for the Access Series CT (6/16/Dual) family. Not confirmed for other Philips CT systems.
Full troubleshooting for ALE19
The complete diagnosis, cause tree, and step-by-step remedy for the Patient Table / Patient Support subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated Patient Table / Patient Support codes
Source: Philips Access CT Service Trouble Shooting Guide — 459800863401 Rev. C. Reference information re-authored by FSELIB — not a copy of the OEM manual.