ALE16 — Patient table servo drive IGBT over-temperature
Philips CT error code, reported by the Patient Table Servo Drive.
What ALE16 means
On Access Series CT (6/16/Dual) CT systems, status code ALE16 indicates: Patient table servo drive IGBT over-temperature. The code originates in the Patient Table / Patient Support section of the Patient Table Servo Drive. Severity level alarm (Alarm — servo drive fault).
Reported action: Let the servo drive cool and check its cooling/ventilation; inspect for excessive duty cycle or an internal fan failure before replacing the drive..
Listed in Table 1-12 'Patient Table Servo Error Codes' of the Philips Access CT Service Trouble Shooting Guide (459800863401 Rev. C), for the Access Series CT (6/16/Dual) family (source text: 'Igbt Temprature Abnormal'). Not confirmed for other Philips CT systems.