ALE14 — Patient table backward limit switch fault
Philips CT error code, reported by the Patient Table Servo Drive.
What ALE14 means
On Access Series CT (6/16/Dual) CT systems, status code ALE14 indicates: Patient table backward limit switch fault. The code originates in the Patient Table / Patient Support section of the Patient Table Servo Drive. Severity level alarm (Alarm — servo drive fault).
Reported action: Check the backward (IN/rear) limit switch and its wiring (see 'IN limite switch LED' / 'Front Limit Switch LED' in Table 1-10/1-11) for a stuck or miswired switch before replacing it..
Listed in Table 1-12 'Patient Table Servo Error Codes' of the Philips Access CT Service Trouble Shooting Guide (459800863401 Rev. C), for the Access Series CT (6/16/Dual) family. Not confirmed for other Philips CT systems.