ALE09 — Patient table servo position control error too large
Philips CT error code, reported by the Patient Table Servo Drive.
What ALE09 means
On Access Series CT (6/16/Dual) CT systems, status code ALE09 indicates: Patient table servo position control error too large. The code originates in the Patient Table / Patient Support section of the Patient Table Servo Drive. Severity level alarm (Alarm — servo drive fault).
Reported action: Check the position feedback (encoder) and mechanical coupling for slippage or binding causing the servo to fall too far behind its commanded position; recalibrate or replace the encoder/drive if needed..
Listed in Table 1-12 'Patient Table Servo Error Codes' of the Philips Access CT Service Trouble Shooting Guide (459800863401 Rev. C), for the Access Series CT (6/16/Dual) family. Not confirmed for other Philips CT systems.