HC10922 — Lower pick-up vacuum valve V2 failure
Agfa CT error code, reported by the Film Tray Pick-up Modules (Sensors, Motors & Actuators).
What HC10922 means
On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC10922 indicates: Lower pick-up vacuum valve V2 failure. The code originates in the Film Tray Pick-up — Sensors, Motors & Actuators section of the Film Tray Pick-up Modules (Sensors, Motors & Actuators). Severity level alarm (Alarm).
Reported action: Check the connections to V2, then connector XS4, then connector P31 on the MEC board, reconnecting or replacing any faulty connector found. Check the wiring from V2 to XS4 and from XS4 to MEC/P31 with a multimeter and repair as needed. If the valve is confirmed defective, power off the printer and replace V2; if the fault traces to the board, replace the MEC board..
Raised on the Drystar 5500/5503 when the lower pick-up module's lower-magazine vacuum valve V2 is not connected or is malfunctioning. Filed in the pick-up module sensors/motors/actuators fault-code table with the equivalent upper-tray vacuum-valve code.