GE Healthcare · CTBrightSpeed260107223260107236

BrightSpeed 4/8/16 slice

Symptom

"Scan hardware Stopped Scan" with Axial Errors. "Scan hardware Stopped Scan" with Axial Errors

Error message

Error: 260107223 Gantry (Axial) could NOT find the 'C' pulse See LD for full error Mon May 2 09:25:53 2005 Error: 260107223 Host: tgp Process: tgp_subsys File: amc_mgr.c Line: 3551 Exception Class : Abort Severity : Primary File : amc_mgr.c Line# : 3551 Function : AXIAL CONTROL : Axial Control Scan Type : Helical Scan: 19/1/1 Gantry (Axial) could NOT find the 'C' pulse . (AMC State:4 Cmd:21 CurVel: 8.800 sec/rev). HW/SW Error: Retry Rx to verify probable Gantry Hardware problem.

Error: 260107236 Axial Motor Control (AMC) Logic will NOT Comand the DRIVE until the Gantry motion is STOPPED. See Long Description for full error Mon May 2 09:25:58 2005 Error: 260107236 Host: tgp Process: tgp_subsys File: amc_mgr.c Line: 2385 Mon May 2 09:25:58 2005 Host : Table Gantry Processor Ermes # : 260107236 Exception Class : Abort Severity : Primary File : amc_mgr.c Line# : 2385 Function : AXIAL CONTROL : Axial Control Scan Type : Helical Scan: 19/1/1 Axial Motor Control (AMC) Logic will NOT Comand the DRIVE until the Gantry motion is STOPPED. The Encoder and the Drive MUST be Sync'd/Referenced. (Cur State: 10 Cmd: 21) HW/SW Error: Retry Rx to verify probable Gantry Hardware problem.

Applies to

Models
BrightSpeed
Product
BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2277225 Axial Brake assembly ( Holding Brake for Axial Drive ) See Long Description for more info 2277225 Axial Brake assembly ( Holding Brake for Axial Drive ) Part #: 2277225 Part Type: GP CONSUMABLES Part Weight: 2.27 LBR Repairable: No Returnable: No Hazardous Material: N Description: BRAKE ASSY, H2 AX DRIVE MOTOR FRU: Yes Status: Orderable Comments: FLS - H2 GANTRY: MUST ORDER FROM HOLT ELECTRONIC. MS26561 DSM 1/31/
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required