Infinia gantry Error #155 it says that there is an ANALOG switch error for whatever axis, what switch in the s…
Infinia gantry Error #155 it says that there is an ANALOG switch error for whatever axis, what switch in the system are we actually looking at as the issue
Symptom
Infinia gantry Error #155 it says that there is an ANALOG switch error for whatever axis, what switch in the system are we actually looking at as the issue
Error message
Error #155 (Error 155) TEST: ANALOG SWITCH FAILURE, AXIS=$ Infinia gantry Error #155 comes as part of the motion Safety tests. The Analog switch is a switch in the SEM. This error could come from: A faulty SEM ( where this switch is defective ..) or From some off-set on the SEM to Elmo driver - command signal, or Some off-set in the ELMO Driver
Applies to
- Models
- Infinia
- Product
- Infinia II Infinia II Acquisition System Infinia HE4 Infinia II Hawkeye 4 & Infinia II GP (Infinia running Windows XP software) Infinia I Infinia I Acquisition System
- Component
- *Gantry Gantry Gantry 5317933 (5318465 SEM) - Infinia Hawkeye 4 SEM board 5317933 (5318465 SEM) - Infinia Hawkeye 4 SEM board March 2010 New SEM P/N 5317933 Infinia II/HE4, Compatible in all systems, EXCEPT Infinia 1 Old Part Number 5175923 Infinia Hawkeye 4 SEM board (PCA500013)
- Area
- Nuclear Medicine
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare Nuclear Medicine issues
- Service Note 43783_1 Feb/08 Infinia/Ventri/Xeleris 1.1 & 2
- After software upgrade FMI40786, the Infinia application does not automatically start on boot up
- I need the product data sheet on catalog E4500AK x-ray in use warning light interface control panel for Infinia
- Infinia Does not recognize collimator is mounted to detector, not reading collimator ID correctly.
- Where do I obtain the Non-Linear Correction tables for a VG/Infinia Hawkeye system if can't find the CD that came with the DAS assembly?
- Hawkeye CT scan Ring (Circular) artifacts in X-ray images appearing in the same radius in all of the images.