GE Healthcare · Nuclear MedicineInfinia
Infinia Does not recognize collimator is mounted to detector, not reading collimator ID correctly.
Symptom
Infinia Does not recognize collimator is mounted to detector, not reading collimator ID correctly.
Error message
ERROR_COLLIMATOR_READ:.:Could not read mounted collimator from scanner
Applies to
- Models
- Infinia
- Product
- Infinia II Infinia II Acquisition System Infinia I Infinia I Acquisition System Infinia HE4 Infinia II Hawkeye 4 & Infinia II GP (Infinia running Windows XP software)
- Component
- 5338523 Two Collimator ID Boards (Kit) 2199382 Collimator ID Board Part replaced by 5338523 . This part is for the new Collimator ID design. Please refer to SN 43837 for instructions If your system is equiped with the old design (2199382) then you will need to order 5342601 - First Time Collimator ID UPgrade Kit ( FMI 40822 ). Note: The screws holding the ID pcb's on the collimator may have been secured with hard or permanent loctite making it very hard to remove the screws The screws that holding the Collimator ID board is defined to be secured by Loctite 222 (that can be untightened later). To open the screws should use the right tool - Phillips screwdriver PH #2. 5342601 First Time Collimator ID UPgrade Kit 5342601 - First Time Collimator ID UPgrade Kit - First Time Collimator ID UPgrade Kit ( FMI 40822 ). *Gantry Gantry Gantry
- Area
- Nuclear Medicine
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare Nuclear Medicine issues
- Service Note 43783_1 Feb/08 Infinia/Ventri/Xeleris 1.1 & 2
- After software upgrade FMI40786, the Infinia application does not automatically start on boot up
- I need the product data sheet on catalog E4500AK x-ray in use warning light interface control panel for Infinia
- Where do I obtain the Non-Linear Correction tables for a VG/Infinia Hawkeye system if can't find the CD that came with the DAS assembly?
- Hawkeye CT scan Ring (Circular) artifacts in X-ray images appearing in the same radius in all of the images.
- Service Tool - NM Phantom Carrying Case part numbers