System is Logging ermes Errors With No Apparent Effect on System.

Bug in Software.

Symptom

System is Logging ermes Errors With No Apparent Effect on System. The "ermes" errors of the type shown below are being logged, filling up the message log. The error messages don't explain what's wrong. There doesn't seem to be any effect on the system otherwise. Mon Feb 11 13:29:11 2008 0 4 lx-mr2 scn msgFile.c 315 The message key -4 was not found. Mon Feb 11 13:29:19 2008 -4 4 lx-mr2 scn PsdError.m 377 Ermes error, key: -4, error: 4009.

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Area
MR

Full cause & step-by-step resolution