MKS 3T Amplifier will not go to ready after several power outages at the site.
Symptom
MKS 3T Amplifier will not go to ready after several power outages at the site. What to Check? When powering the amplifier on you get a fault indication and then the amp goes to an off state.
Error message
2254253 3T RF Amplifier PA Filament under Voltage (fault 38). The PA Filament supply output was below specification(15V - 5%). 2254253 NSP SCP:../rf_nb_contro 1460 memp 8HRBRAIN protocol:DQA_8ch_Brain/1 3T RF Amplifier PA Filament under Voltage (fault 38). The PA Filament supply output was below specification(15V - 5%). Possible Causes: Defective supply or short on supply output. If the fault persists, replace the RF amplifier
Applies to
- Models
- Excite
- Product
- 3T Excite HDx 14.0 Software. The software that this solution is tied to is indicated by the Software concept below.
- Software
- 14.0 M5 Excite HDx HDi Platforms only.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.