Pular para o conteúdo principal
GE Healthcare · MR

OrthoMRI Coil locking device is broken

Symptom

OrthoMRI The coil moves even after the locking mechanism is locked Inside the gantry the knee coil is locked by the coil lock mechanism, sometimes also if the coil lock mechanism is open the coil is not perfectly locked: it’s not possible to remove it but it’s possible to rotate it. In this case, it’s necessary to place one or two plastic disks under the coil. Inside the gantry the knee coil is locked by the coil lock mechanism, sometimes if the coil lock mechanism is open the coil is not perfectly locked: it’s not possible to remove it but it’s possible to rotate it. In this case, it’s necessary to place one or two plastic disks under the coil.

Applies to

Product
OrthoMRI EscanNE. OrthoMRI EscanNE.
Software
OrthoMRI S/W Release 9.1B. OrthoMRI S/W Release 9.1B.
Component
OrthoMRI P/N LNR9101821001 Swing Arm assembly OrthoMRI P/N LNR9102029000 Coil lock assembly- green, blue is not available
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues