Infinia CoDe QC fails FWHM - Energy resolution is over limit When perform FDG CoDe QC, found Energy Peak is ab…

Infinia CoDe QC fails FWHM - Energy resolution is over limit

Symptom

Infinia CoDe QC fails FWHM - Energy resolution is over limit When perform FDG CoDe QC, found Energy Peak is about 511, but the Energy FWHM is more than 11% , like H1 is about 13%, H2 is about 18%.

Applies to

Models
Infinia
Product
Infinia II Infinia II Acquisition System
Component
*CoDe (Coincidence) *CoDe (Coincidence) *CoDe (Coincidence) Option
Area
Nuclear Medicine

Full cause & step-by-step resolution