GE Healthcare · CTHiSpeed
HiSpeed QX/i - Octane
Symptom
System will not come up to applications with image space errors. System will not come up to applications with image space errors
Error message
Error: 9043 opendir failed : /usr/g/sdc_image_pool2/reserved See long description for full error Thu Mar 6 05:03:45 2008 Error: 9043 Host: mrep_ct01 Process: imserver File: diskmisc.c Line: 426 opendir failed : /usr/g/sdc_image_pool2/reserved /usr/var/adm/SYSLOG: ERROR: Unable to mount /usr/g/sdc_image_pool2 usr/var/adm/SYSLOG: ERROR: Unable to mount /usr/g/sdc_image_pool2
Applies to
- Models
- HiSpeed
- Product
- HiSpeed QX/i - Octane Always MDAS4 101I.2_HSQXIM5 101.4_HSQXIM4 HiSpeed QX/i - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 101I.2_HSQXIM5 101.4_HSQXIM4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2169940-81 Octane Computer 73 GB Disk Drive - replaces 2169940-78 (18GB) See long description for aditional information 2169940-81 Octane 73 GB Disk Drive, replaces 18 GB Disk 2169940-78 Part #: 2169940-81 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: CT OCTANE 2 73 GB DISK DRIVE FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.