GE Healthcare · CTLightSpeed2601151917156429999
Scanning aborted during a scout scan with Cradle position is out of tolerance.
Symptom
Scanning aborted during a scout scan with Cradle position is out of tolerance.
Error message
Error: 260115191 Cradle position is out of tolerance See long description for full error Thu Sep 1 09:04:12 2005 Error: 260115191 Host: tgp Process: tgp_subsys File: mtcb_cradle_monitor.c Line: 414 Thu Sep 1 09:04:12 2005 Host : Table Sub-System Ermes # : 260115191 Exception Class : Soft Severity : Pri/Soft File : mtcb_cradle_monitor.c Line# : 414 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Cradle position is out of tolerance. cradle_position(um) : 1666 target_position(um) : 71564 velocity (um/s) : 29999 fpga status : 0x8140
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2339120 Cradle drive assembly See long decription for details Part #: 2339120 Part Type: GP REPAIRABLES Part Weight: 2.20 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: CRADLE DRIVE ASSY - H PWR FRU: Yes Status: Orderable Comments: FLS - CRADLE DRIVE ASSY - H PWR. ADDED PER HPOWER PROGRAM RSPL BY MS20544 ON 2/24/03. POSSIBLE M3 DATE: FW19, 2003.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.