GE Healthcare · CTLightSpeed260003116
Scanner Hardware Stopped Scan with error 260003116
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260003116 Filter drive tracking detected no motion when motion was requested. See long description for sample error. Fri Sep 14 10:15:52 2007 Error: 260003116 Host: orp Process: orp subsys File: ccb drive timers.c Line: 504 Fri Sep 14 10:15:51 2007 Host : Collimator Ermes # : 260003116 Exception Class : Abort Severity : Primary File : ccb drive timers.c Line# : 504 Function : No System Function Reported Scan Type : Helical Scan: 8516/3/20 Filter drive tracking detected no motion when motion was requested. Verify operation of stepper motor and encoder feedback Start Position = 41999um 39697cnt Desired Position = 0um 0cnt Current Motor Position = 40415um 38200cnt Current Encoder Position= 41998um 39696cnt
Applies to
- Models
- LightSpeed
- Product
- LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5140001-4 helios 4 collimator with notched filler bar and rating plate used on 5.X scanners. See Long Description for Details. Part #: 5140001-4 Part Type: GP REPAIRABLES Part Weight: 2.20 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: helios 4 collimator with notched filler bar and rating plate FRU: Yes Status: Orderable Comments: Added by ECR 2013518 as ms13452 050428 **FE Assist** Used on H-power - main difference is the mount holes are larger to accommodate for larger bolts to hold the added weight of the Hercules Tube. 5140001-4 is also used on HP-60 s . 514001-4 is not backwards compatible.(12/5/06 212001261) *** 2ND OPINION NOTES *** Replacing for artifacts? Look into collimator with bright light. Any particulate? Look at plastic over copper filter, any distortion or yellowing? Bow tie filter ok? There are several lower level FRUs. Ensure that one of the lower level FRUs could not fix the problem.
- Area
- CT