GE Healthcare · CTLightSpeed260111007225673032636
Gantry display blank with error 260111007
Symptom
Gantry Display blank. Gantry Display blank
Error message
Error: 260111007 Communication reestablished to GUIF CAN node ID# 1 (or nodes 2,3,4) See LD for details Wed Apr 4 16:55:13 2007 Error: 260111007 Host: tgp Process: tgp_subsys File: guif_can_mgr2.c Line: 1908 Wed Apr 4 16:55:13 2007 Host : Table Gantry Processor Ermes # : 260111007 Exception Class : Secondary Severity : Pri/Soft File : guif_can_mgr2.c Line# : 1908 Function : OPERATOR INPUT/OUTPUT : Gantry UIF Communications Scan Type : None/Unknown Scan: 9106/1/0 Communication reestablished to GUIF CAN node ID# 1. Node part number 2256730. Node serial number 32636.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2256730-3 Display assembly used on 2.X/3X/4X5X/7X See long description for details. Part #: 2256730-3 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: H-POWER COMPATIBLE GANTRY DISPLAY FRU: Yes Status: Orderable Comments: FLS - H-POWER COMPATIBLE GANTRY U/O CT H POWER DISPLAY
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required