GE Healthcare · MRExcite2252306

Cannot scan in Head Transmit Mode due to TR Driver Faults due to LEGACY HEAD COIL CABLE W/COIL ID.

Symptom

Cannot scan in Head Transmit Mode with Head TR Driver Faults. Body Transmit works OK.

Error message

2252306 TR Driver Fault Detected Short Circuit Source of Fault is Head TR See Long Description for Extended Error. 2252306 ERROR: EM_DCB_TR_SHORT_CKT TR Driver Fault Detected - Short Circuit. Source of Fault is %b. Value: %d mA Limit: %d mA Extended Error: If this error happens repeatedly then check for a short-circuited PIN diode or other short-circuit condition somewhere in the failing transmit chain. Use TR DD MC System Path diagnostic or Gather Load Data and verify that the short circuit values are less than 30mA. If error happens intermittently and cannot be traced to failed hardware then it may be due to gradient noise in the system.

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2384268 CLASSIC HEAD COIL CBL ASM W/COIL ID See Long Description and Attachment See Attachment for the engineering drawing of this part: 2384268ADW Rev. 2. This is not available online. To confirm this is the most appropriate drawing, go to ePDM.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.