Pular para o conteúdo principal
GE Healthcare · MRExcite20000235222521062000023722252109

Applications Software Will Not Boot Because The System Failed To Determine Hardware State For AGP Board Within 20 Second…

Applications Software Will Not Boot Because The System Failed To Determine Hardware State For AGP Board Within 20 Seconds.

Symptom

HP Linux Workstation Fails To Boot, Hangs At Do Not Interact Message

Error message

200002352 A HOST_SHUTDOWN was initiated by the death of NSP.

2252106 Diagnostic software error. Failed to determine hardware state for AGP, SCP board. The board may be busy loading vxWorks or may be hung. Also check the serial cable for the board and the serial terminal server. 2252106 Diagnostic software error. Failed to determine hardware state for AGP, SCP board. The board may be busy loading vxWorks or may be hung. Also check the serial cable for the board and the serial terminal server.

200002372 The process NSP took too long to start up. Thu Jan 27 08:56:01 2005 Error: 200002372 Host: lx-mr Process: hostMonitor File: ssProcStart.c Line: 933 The process NSP took too long to start up.

2252109 Power up test found the following errors: APS power-up diagnostics failed RAM/ADR test Mon Nov 14 06:51:57 2005 Error: 2252109 Host: lx-sr Process: NSP File: pwrup test.c Line: 145 Power up test found the following errors: APS power-up diagnostics failed RAM/ADR test

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2294300-3 MGD CHASSIS AGP BOARD NOTE> this part is misspelled in FEMC as APG
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.