Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT230011067

ID: 98087 LightSpeed 4.X

Symptom

Scanner Hardware Shutdowns and Recon problems Lots of error messages dealing with Scan Disk Array, missing images, deleted and removed header files.

Error message

Error: 230011067 Unexpected file, or directory, detected in scan data root directory on scan data disks during start-up. The unexpected file, or directory,will be removed. See LD for complete error Error: 230011067 Unexpected file, or directory, detected in scan data root directory on scan data disks during start-up. The unexpected file, or directory,will be removed. Fri Feb 10 05:49:23 2006 Error: 230011067 Host: ct01 Process: scanfilemgr File: sdio xfs api.c Line: 3500 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Unexpected file, or directory, detected in scan data root directory on scan data disks during start-up. The unexpected file, or directory, will be removed. File/directory name: lost+found Root directory: /raw data

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues