Independent Vector Gating unavailable in scan application due to PAC Type is set incorrectly.
Symptom
Independent Vector Gating unavailable in scan application. "Independent Vector Gating" is the new Cardiac Gating option starting with Excite HD 12.0 software. But the technologist is unable to select it in the Scan Rx. It is unavailable.
Error message
2226158 The system detected a PAC II connected. Please check the PAC version connected to the system. Vector Gating will ONLY operate reliably with PAC III version.
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution