Pular para o conteúdo principal
GE Healthcare · MRExcite2254265

2254265 3T RF Amplifier PA Bias Current fault 61 due to 3T 35 KW RF deck Top Half.

Symptom

Unable to Scan with 3T MKS RF Amplifier will Not go Ready MKS amp powers up but will not come ready.

Error message

2254265 3T RF Amplifier PA Bias Current fault 61 See Long Description Error: 2254265 Host: fmri3tw Process: NSP File: SCP:../RfNbMdsHWO.cpp Line: 2218 3T RF Amplifier PA Bias Current (fault 61). Possible Causes: defective PA tube, defective supply.

Applies to

Models
Excite
Product
3T Excite HD 12.0 software only
Software
12.0 Excite HD and 3T Excite HD products only
Component
2336064-14 35KW RF deck Top Half
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.