Scan room Door is Open Errors Due to IRF I/O board.

Symptom

Auto and Manual Prescan Fails

Error message

2248631 scan room door is open Mon Jan 7 15:02:15 2008 Error: 2248631 Host: mr01 Process: NSP File: APS:aps_cf Line: 238 Exception Class: Unknown Severity: Unknown Function: Software/Firmware/Config: N/A scan room door is open

2224666 Auto Prescan failed Scan room door is open 2224666 Host: lx-mr Process: psc File: PscMainControl.cc Line: 1322 Auto Prescan failed, Scan room door is open.

Applies to

Models
Excite
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 14.0 For Hino Excite HDe 14.0 systems only.
Component
2274196 IRF I/O Board See Long Description. Per FEMC part information, Description of part number 2274196 is an INTERFACE and REMOTE RF TRANSITION BOARD. Part Comments as of 2/8/2008 are as follows; FLS - INTERFACE & REMOTE RF TRANSITION ADDED PER RSPL. USED ON PHOENIX MGD. MSA8429 05/14/02 ROUTE TO MR FOR CONSUMABLE COLLECTION PER B. HAHLEN, 7-11-02, A. MASCH THIS PART WAS ADDED AS PART OFTHE EXCITE NPI FOR MR. 9/01/02 MS24445 BH FLS - FE ASSIST 08OCT2002 (MS2045) ADDED TO MR DISTRIBUTION LIST FOR PARTS TRACKING.