GE Healthcare · MRExcite2252305
TR Driver Fault Detected Open Circuit Source of Fault is Body TR
Symptom
Unable to Scan Body with TR Driver Faults Unable to scan body with TR Driver faults
Error message
2252305 TR Driver Fault Detected Open Circuit Source of Fault is Body TR. Thu Aug 5 11:00:38 2004 Error: 2252305 Host: fmri3te Process: NSP File: SCP:DCB Line: 1197 TR Driver Fault Detected - Open Circuit. Source of Fault is Body TR. Value: 16 mA Limit: 2400 mA->
Applies to
- Models
- Excite
- Product
- 3T Excite G3 Software ONLY
- Software
- G3 For 3T Excite only
- Component
- 2370221 3T DUAL BODY HYBRID aka 3T DUAL BODY T/R SWITCH ASM See Long Description In Renewal Parts, the part name is "3T Dual Body Hybrid". In FEMC the same part number has the part name "3T DUAL BODY T/R SWITCH ASM". Part # 2370221. This part is for the LCC300 short bore 3T both on G3 systems and Excite HD/HDx/HDxT. NOT compatible with the 3T94 long bore 3T.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.