GE Healthcare · CTLightSpeed184845320471822162836080327170300
Cradle will not drive in or out
Symptom
Cradle will not drive in or out. Cradle will not drive in or out
Error message
Error: 184845 Cradle unable to move to position 32047, internal cradle error status 1500 see Long Description for full error Fri Sep 1 15:04:42 2006 Error: 184845 Host: ETC Process: Cradle File: cmc interface.c 1.30 9/25/96 Line: 577 Function: No System Function Reported Scan: 1145/2/5 Type: Axial Exception Level: Pri/Most Severe Time: 15:04:41:730 Log Series: 453 Cradle unable to move to position 32047, internal cradle error status 1500 Error: 182216 HW/SW error: Motion was prohibited by hardware Limit switches. Hardware Inhibit Switches 0x84dc See long descripton for full sample error message. Fri Dec 21 10:12:18 2007 Error: 182216 Host: ETC Process: Cradle File: cmcStart.c 1.60 7/23/96 Line: 415 Function: PATIENT ANATOMY POSITIONING : Cradle Drive Scan: 28360/2/2 Type: Helical Exception Level: Pri/Soft Time: 10:12:19:245 Log Series: 408 HW/SW error: Motion was prohibited by hardware Limit switches. Hardware Inhibit Switches 0x84dc Current Position 80327 (.01 mm) Dir: 1 (0 = Out 1 = In [Gantry]) Characterization Limits 0 (.01 mm) 170300 (.01 mm) Error: 182211 Velocity error detected by hardware (LM628 excessive position error). See the Long Description for the full error Tue Au
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2269175-3 LONGITUDINAL ENCODER ASM (Used on HSA, CTi, H1 QXi, Plus, Ultra, 16 slice H4 and HiSpeed QXi.) replaces 2269175-2. See long description for details Part #: 2269175-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Longitudinal Encoder Assembly with Dynapar encoder FRU: Yes Status: Orderable Comments: Per ECR 2036924, 212008898, 12/4/2006. ** FE Assist ** If no stock or unable to order the 2269175-3 , order 2269175-2. (212001261 5/22/07) Part #: 2269175-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Longitudinal Encoder Asm. NOTE:This asm. has LCC encoder/tested and sensor gap is 0.06-0.08mm FRU: Yes Status: Orderable Comments: Per ECR 2036924, 212008898, 12/4/2006
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required