Pular para o conteúdo principal
GE Healthcare · MRExciteOvation2244592

CWA ID97634. Will Not Scan with Gradient Errors.

Symptom

Will Not Scan with Gradient Errors.

Error message

2244592 SGA Power Supply Under Voltage Fault 2244592 "SGA Power Supply Under Voltage Fault.\n\ Look at SGA-PS front LEDs. If BUS_OV LED lit, check PDU\n\ 420V output. If UV_X,Y,Z LED lit, swap military connectors\n\ to isolate SGA versus SGA-PS."

Applies to

Models
Excite, Ovation
Product
3T Excite HD 12.0 software only
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR

Full cause & step-by-step resolution