Pular para o conteúdo principal
GE Healthcare · MRSigna22255392225536

Cannot Scan Due to SSM Calibration.

Symptom

Cannot Scan With intermittent TR and Dynamic Disable Faults. The customer when scanning is getting intermittent TR and Dynamic Disable faults. It happens with all coils and cannot be pinned down to a specific PSD.

Error message

2225539 Dynamic disable fault, Front of Coil open. Dynamic Disable bias fault on front drivers. Wed Dec 1 09:09:29 2004 Error: 2225539 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_tr_mon Line: 6140 Dynamic disable fault, Front of Coil open. Dynamic Disable bias fault on front drivers. Possibly a disconnected cable between the driver board and the dynamic disable switch boxes, or an open diode in dynamic disable switch boxes on FRONT of BRM.

2225536 Dynamic disable and T/R driver board error. A spurious dynamic disable fault occurred. See Long Description Error: 2225536 Suite: MRS1 Host: OC0 Proc: HPC_PORT_ID PMH.m 1.29 Method: spi_tr_mon Line: 6220 Dynamic disable and T/R driver board error. A spurious dynamic disable fault occurred. It cleared itself before the specific problem could be read. Possible causes: noise on the dynamic disable board or a bad board.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required