GE Healthcare · CTLightSpeed182408
LightSpeed 3.X (Ultra, Plus, QXi) - Octane
Symptom
Table elevation moves slowly with a large patient. Table elevation moves slowly with a large patient
Error message
Error: 182408 Velocity tracking error. Probable Problem could be 180V Power is LOW. See Long Description for complete error Error: 182408 Host: ETC Process: Elevation File: CmdMotion.c 1.35 12/23/93 Line: 83 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 11:57:28:685 Log Series: 35 HW/SW error: Velocity tracking error. Actual Velocity 0 (.1 mm/sec) Expected Velocity 400 (.1 mm/sec) Tol 320 (.1 mm/sec) Cur Pos 3276 (.1 mm) Elev Direction: 1 (0 = Up 1 = Down). INFO: Probable Problem could be 180V Power is LOW.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2100671 CT TABLE GAS SPRING - Used on ALL HSA, CTi, LightSpeed 1.X through 5.X tables Part #: 2100671 Part Type: GP CONSUMABLES Part Weight: 1.40 LBR Repairable: No Returnable: No Hazardous Material: N Description: CT Table gas spring FRU: Yes Status: Orderable Comments : FLS - REPLACES 46-327064P1 & 46-296891P1 PER PCN 161843 7-20-93 MS4398. NOT ON SSS: SUSPA IS NOT SUPPLIER. TOOL FOR REPLACING GAS SPRING IS 2144261 ************************************************************PART HAS BEEN DE-LISTED AS A HAZARDOUS MATERIAL ITEM. PART IS NOT HAZARDOUS FOR ANY MODE OF TRANSPORTATION. HAZARDOUS PAPERWORK IS NOT NEEDED. 4/99 ************************************************************ ALSO USED ON QXI LIGHTSPEED ALSO USED ON LIGHTSPEED GELS - "NON FLAMMABLE GAS - HAZ IATA DESC. ""ARTICLES, PRESSURIZED, PNEUMATIC "" UN 3164 - CLASS 2.2",GAS SPRING HAZ #8 is the shock
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required