Pular para o conteúdo principal
GE Healthcare · MRSigna2225552

Unable to Scan with Multi Coil TR Driver Faults

Defective coil.

Symptom

Unable to Scan with Multi Coil TR Driver Faults

Error message

2225552 Multi-Coil T/R Driver Board Fault. An Open Circuit was Detected on Port 2. Thu Aug 21 12:06:10 1997 Suite: MRS1 Host: OC0 Proc: HPC_PORT_ID Error: 2225552 File: PMH.m 1.29 Method: spi_tr_mon Line: 6240 Multi-Coil T/R Driver Board Fault An open circuit was detected on port 2. Check cable to multi-coil board or check for an open pin diode.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2225545-6 USAI PREMIER 7000 PHASED ARRAY CTL SPINE COIL. M1087SP No special tuning is required after replacing the Adapter Block 2225545-9 of the M1087SP USAI PREMIER 7000 PHASED ARRAY CTL SPINE COIL 2225545-6.
Area
MR