Pular para o conteúdo principal
GE Healthcare · Nuclear MedicineInfinia4199424401400

Infinia Querying the studies from a HIS/RIS server (MWL) returns no entries, although all the MWL settings hav…

Infinia Querying the studies from a HIS/RIS server (MWL) returns no entries, although all the MWL settings have been correctly configured

Symptom

Infinia Querying the studies from a HIS/RIS server (MWL) returns no entries, although all the MWL settings have been correctly configured Infinia Querying the studies from a HIS/RIS server (MWL) returns no entries, although all the MWL settings have been correctly configured. Modality Worklist (MWL) query returns no entries due to DICOM tag [Reason for Study] (0032,1030) containing more than 64 characters - PQRs 13087532 and 13143771

Error message

MC_Get_Value_To_String failed Value too large for buffer ERROR=Error in NetworkServer d:\tacq\cu\src\cxx\dm\nw2\mergeobj.cpp(72)::MC_Get_Value_To_String failed Value too large for buffer The following error is logged in the NetworkServer log (D:\tacqdb\log\[email protected]): ERROR=Error in NetworkServer d:\tacq\cu\src\cxx\dm\nw2\mergeobj.cpp(72):: MC_Get_Value_To_String failed Value too large for buffer

buildWorkListLevelResponseList:: couldn't get value for tag 4199424 WARNING=Warning in NetworkServer d:\tacq\cu\src\cxx\dm\nw2\dicomserviceuser.cpp(1425) buildWorkListLevelResponseList:: couldn't get value for tag 4199424 The following warning is logged in the NetworkServer log (D:\tacqdb\log\[email protected]): WARNING=Warning in NetworkServer d:\tacq\cu\src\cxx\dm\nw2\dicomserviceuser.cpp(1425) buildWorkListLevelResponseList:: couldn't get value for tag 4199424 NOTE: The DICOM tag in this warning message is displayed as a decimal value. DICOM tags are typically displayed as hexadecimal values. It is necessary to convert the decimal value to hexadecimal to properly interpret the DICOM tag that is causing the issue. In this case decimal 4199424 converts to hexa

Applies to

Models
Infinia
Product
Infinia I Infinia I Acquisition System
Software
Infinia 1.004.009.21 Infinia M4 version is 1.004.009.21 Infinia M4 version is 1.004.009.21
Component
*Software Software Software
Area
Nuclear Medicine

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.