GE Healthcare · CTLightSpeed183365501681831825000144858
LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 183365 Cannot bring HVDC bus up for xx seconds See long Description for complete error Wed Apr 5 06:52:02 2006 Error: 183365 Host: OBC Process: Generator File: gen_rotor.c 1.39 Line: 320 Function: XRAY GENERATION : Tube Rotor Control Scan: 50168/5/8 Type: Static Exception Level: Pri/Most Severe Time: 06:52:02:815 Log Series: 118 Cannot bring HVDC bus up for 6 seconds. Error: 183182 I/O board pilot relays are closed. PDU pilot relay not energized. HVDC Rail Voltage is xxxx See Long Description for more details. Error: 183182 I/O Bd. pilot relays are closed. PDU pilot relay not energized. HVDC is xxxx. ** Note the HVDC Rail Voltage. It is important in troubleshooting Tue Sep 21 06:13:59 2004 Error: 183182 Host: OBC Process: Generator Function: SYSTEM POWER CONTROL: HVDC Control Scan: 50001/1/1 Type: Axial Exception Level: Pri/Most Severe I/O board pilot relays are closed. PDU pilot relay not energized BCSTAT = 0 HVDC Rail voltage = 0V Sun Dec 12 18:11:57 2004 Error: 183182 Host: OBC Process: Generator File: gen_rotor.c 1.39 Line: 488 Function: SYSTEM POWER CONTROL : HVDC Control Scan: 44858/1/1 Type: Scout Exception Level: Pri/Most Severe Time: 18:11:57:355 Log Series: 807 I/
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2 This is HP CPU host running LINUX with Pegasus IG. 308L.8_H3.3M4 ~ 4 or 8 slice MDAS. ~ Could be an Ultra, Plus or QXi system. Must run swokinstall -p to verify. ~ See Long description for details . LightSpeed 3.X MDAS4 Gantry and GOC2 Gantry - 2281177-2, DAS - 2272768, Console – 2341104-2, Table - 2276776, PDU - 2269902 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2133533-3 CAPACITOR, 4600UF, 450V, 14A, 85DEGC Quantity 2 required See Long Description for details 2133533-3 Capacitor, 4600 uf, 450v, 85 deg, Quantity 2 required and replace both if you suspect either one has failed or causing problems. Part #: 2133533-3 Part Type: GP CONSUMABLES Part Weight: 2.27 LBR Repairable: No Returnable: No Hazardous Material: N Description: CAPACITOR, 4600UF, 450V, 14A, 85DEGC FRU: Yes Status: Orderable Parts Offered as: Current: 2133533-3 Harvested: Not Available Exchange: Not Available Comments: FLS - 4600DF, 450V, 14A, 85 DEGREE CU/O COMPACT PDU RHAPSODE CAN SOURCE FROM ON-LINE. MS4636 11/7/96
- Area
- CT
Full cause & step-by-step resolution