GE Healthcare · CTLightSpeed26000240350241
LightSpeed 2.X (Plus, QX/i) - Octane
Symptom
"Scan hardware Stopped Scan" with DAS power errors. "Scan hardware Stopped Scan" with DAS power errors
Error message
Error: 260002403 The Data Acquisition System Has Detected An Error In The +5V Converter Board Digital DAS Power Supply. See long descripton for a sample of complete error message. Mon Dec 10 01:59:42 2007 Error: 260002403 Host: DAS Process: DCB Firmware File: pwrtst.c Line: 674 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 50241/1/0 Type: None/Unknown Exception Level: Sec/Soft Time: 01:59:42:620 Log Series: 208 The Data Acquisition System Has Detected An Error In The +5V Converter Board Digital DAS Power Supply.^B Measured: 4749 mV, Expected range: 4750 to 5250 mV
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2225212-3 DAS 5 volt logic power supply Used on SDAS and MDAS See long description for more information. Part #: 2225212-3 Part Type: GP CONSUMABLES Part Weight: 9.07 LBR Repairable: No Returnable: No Hazardous Material: N Description: PWR SUPPLY 5VDC ASM, F-CASE, W/O PLT FRU: Yes Status: Orderable Comments: 5V 35 AMP LOGIC POWER SUPPLY. USED ON WARP 3.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.