Pular para o conteúdo principal
GE Healthcare · MRExcite

HIS RIS Worklist Comes Up Empty Due to Corrupt Configuration Files.

Symptom

HIS RIS Worklist Comes Up Empty

Error message

Error code 4 WLnewPatientId Could Not Get valid Worklist Error seen in /usr/g/hisris/wlsErrors file: Wed 02-Feb-05 06:26:48 WLscanCallbacks.c (1710) Error code: 4 WLnewPatientId(): Could Not Get valid Worklist Error seen in /usr/g/hisris/wlsErrors file: cd /usr/g/hisris more wlsErrors Wed 02-Feb-05 06:26:48 WLscanCallbacks.c (1710) Error code: 4 WLnewPatientId(): Could Not Get valid Worklist

Error Code 17 DICOM server not started at the given port number Thu 07-Feb-08 15:14:21 WLServer.c (1580) Error code: 17 DICOM server not started at the given port number

Error Code 28 Fatal Unknown DCM ERROR Thu 07-Feb-08 15:14:21 WLServer.c (311) Error code: 28 Fatal Unknown DCM ERROR

No Worklist Found on the Remote HIS for the Input Filter See Long Description As seen in the error file /usr/g/hisris/wlsErrors. more /usr/g/hisris/wlsErrors Thu 17-Apr-:3 13:17:10 WLServer.c (1735) Error code: 25 No Worklist found on the remote HIS for the input filter

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Area
MR