GE Healthcare · MRExcite2254253

Unable to Scan with 3T MKS RF Amplifier will Not go Ready.

Symptom

Unable to Scan with 3T MKS RF Amplifier will Not go Ready MKS amp powers up but will not come ready.

Error message

2254253 3T RF Amplifier PA Filament under Voltage (fault 38). The PA Filament supply output was below specification(15V - 5%). 2254253 NSP SCP:../rf_nb_contro 1460 memp 8HRBRAIN protocol:DQA_8ch_Brain/1 3T RF Amplifier PA Filament under Voltage (fault 38). The PA Filament supply output was below specification(15V - 5%). Possible Causes: Defective supply or short on supply output. If the fault persists, replace the RF amplifier

Applies to

Models
Excite
Product
3T Excite HDx 14.0 Software. The software that this solution is tied to is indicated by the Software concept below.
Software
14.0 For both Excite HDx and 3T Excite HDx systems.
Area
MR

Full cause & step-by-step resolution