Scanner hardware stopped scan with KV errors :Initialize and reset TNT Data

Symptom

Scanner hardware stopped scan with KV errors. Scanner hardware stopped scan with Error: 260118215 (60-0318)kV did not reach 75 percent after 20ms.

Error message

Error: 260118215 (60-0318) kV did not reach 75 percent after 20ms See Long Description for full error Sat Mar 25 15:35:28 2006 Error: 260118215 Host: orp Process: orp_subsys File: gen_exec.c Line: 2150 Sat Mar 25 15:35:27 2006 Host : Jedi Ermes # : 260118215 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2150 Function : No System Function Reported Scan Type : Static Scan: 50368/1/3 (60-0318)kV did not reach 75 percent after 20ms. Indicates that there was no HV ON signal generated for exposure time count-up.

Applies to

Models
LightSpeed, VCT, BrightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT