GE Healthcare · CT2300190022600050031818850152
Lightspeed 1.X - Octane
Symptom
Images would not reconstruct.
Error message
Error: 230019002 The scan disk I/O layer failed to consume a response from the scan disk server. See LD for complete error Tue Nov 2 03:19:30 2004 Error: 230019002 Host: sgredct1 Process:scanfilemgr File: sdio proxy.c Line:174 The scan disk I/O layer failed to consume a response from the scan disk server. function:consume sequence:6708 Error: 260005003 ICE CPU did not respond to polling within 180 seconds. Recon subsystem is presumed to be down. Some images may have been suspended. Thu Apr 29 06:05:11 2004 Error: 260005003 Host: fitzipct Process: procICE File: iceCpu.cxx Line: 343 ICE CPU did not respond to polling within 180 seconds. Recon subsystem is presumed to be down. Some images may have been suspended. Last ping was at Sun May 30 22:21:46 2032 with seq num 1818850152 iceConsole.log: Initializing System Memory (DRAM)... System Memory: 128MB, ECC Enabled (ECC-Memory Detected) PPC1-Bug SDC_IG_INTF: >> in the iceConsole.log See Long Description for complete error. The following errors were found in the iceConsole.log cd /usr/g/service/log tail -1000 iceConsole.log | mor e failed to open config file ig_reconif_dbg.cfg S_nfsLib_NFSERR_NOENT SDC_IG_INTF: <<<aborting...>>> restor
Applies to
- Product
- Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2118899-2 RHAPSODE IMAGE GENERATOR Board (IG Board). Used in Lightspeed 1x, CTi 4x, CTi 6x systems. See LD for details Part #: 2118899-2 Part Type: GP REPAIRABLES Part Weight: 3.64 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: RHAPSODE IMAGE GENERATOR FRU: Yes Status: Orderable Comments: FLS - IMAGE GENERATOR BOARD USED ON RHAPSODE MS1356 BURR 12-95 REPLACED 2118899 PER PCN 189598. MS4636 2/6/96 PCN 189939 GELS - KANBAN CSS F00,PCB IMAGE GENERATOR
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.