GE Healthcare · CTLightSpeed18241039139
System exhibiting Error: 182410 HW/SW error: Motion was prohibited by hardware Limit switches
Symptom
Table will not move, gantry will not reset , Elevation No Index Pulse between Interference switches.
Error message
Error: 182410 HW/SW error: Motion was prohibited by hardware Limit switches Thu Aug 2 21:40:36 2007 Error: 182410 Host: ETC Process: Elevation File: emcControl.c 1.23 11/13/92 Line: 168 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 39139/2/0 Type: None/Unknown Exception Level: Pri/Soft Time: 21:40:37:175 Log Series: 764 HW/SW error: Motion was prohibited by hardware Limit switches. Hardware Inhibit Switches 0xc2d8 Current Position 4943 (.1 mm) Dir: 1 (0 = Up 1 = Down). Characterization Limits 5005 (.1 mm) 250 (.1 mm)
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT