GE Healthcare · CTLightSpeedVCT2601210212050020316
Gantry Tilt moves a small distance
Symptom
Gantry tilt moves a small distance then stops (less than 2 deg) Gantry tilts forward a small distance then stops
Error message
Error: 260121021 Tilt Stopping Error. Tilt did not reach target angle See LD for complete error Fri Dec 21 16:47:02 2007 Error: 260121021 Host: tgp Process: tgp_subsys File: tilt_monitor.c Line: 299 Fri Dec 21 16:47:01 2007 Host : Table Gantry Processor Ermes # : 260121021 Exception Class : Secondary Severity : Pri/Soft File : tilt_monitor.c Line# : 299 Function : No System Function Reported Scan Type : Helical Scan: 7649/2/1 Tilt Stopping Error. Tilt did not reach target angle -20500 mdeg. Tilt current angle: -20316 mdeg. Tilt angle tolerence: 250 mdeg. Timeout: 1000 msec.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required