Pular para o conteúdo principal
GE Healthcare · CTLightSpeed244584178570

Unable to scan with Collimator/DAS reset errors

Symptom

Unable to scan with a message on screen that says: Collimator/DAS reset failed.

Error message

Error: 244584 Collimator/DAS reset failed See Long Description for complete error Error: 244584 Host: qx-ct Process: fwmgr File: HwManager.c Line: 1536 Collimator/DAS reset failed.

Error: 178570 Coll/Das controller reports failure in initializing tasks See Long Description for complete error Mon Mar 28 03:39:48 2005 Error: 178570 Host: ct04 Process: fwmgr File: HwSetUp.c Line: 3178 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 OBC_SS timeout: Coll/Das controller reports failure in initializing tasks.

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GOC2 405L.8_H4.2M4 LightSpeed 4.X with GOC2 and MDAS 16 Gantry H4 Gantry Model # 2339985 and Global Operator Console 2 Model #'s 2366438 and 2366438-2(Hino) with HP8000 CPU (or newer) and Pegasus Recon ICE Box Product: LightSpeed16 405L.8_H4.2M4 CPU: GenuineIntel Intel(R) Xeon(TM) CPU 2.66GHz 4 2657 MHz i686 Processors Main memory size: 2048 Mbytes [root@presbyct2]# sg_map -i /dev/sg0 /dev/sda SEAGATE ST336753LW HPS2 /dev/sg1 /dev/sdb SEAGATE ST373453LW HPS1 /dev/sg2 /dev/sdc SONY SMO-F551-SD 1.05 /dev/sg3 /dev/scd0 MATSHITA DVD-RAM SW-9574S AZX8 /dev/sg4 /dev/scd1 HL-DT-ST DVD-ROM GDR8163B 0B15 [root@presbyct2]# df -h Filesystem Size Used Avail Use% Mounted on /dev/sda2 4.0G 1.7G 2.3G 41% / /dev/sda1 27M 12M 14M 44% /boot none 1008M 0 1007M 0% /dev/shm /dev/sda6 14G 3.5G 10G 25% /usr/g /dev/sda7 9.8G 753M 9.0G 8% /usr/g/sdc_image_pool /dev/sdb1 68G 28G 40G 41% /usr/g/sdc_image_pool2 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5191271 MDAS 16 Fuse Holder Upgrade Kit. Supercedes part number 46-170042P28. See Long Description for more information. Part #: 5191271 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: MDAS 16 Fuse Holder Upgrade Kit FRU: Yes Status: Orderable Comments: Per ECR 2036788, 212008898, 11/30/2006. MDAS Fuse Holder Replacement Kit Contents: Part #: 46-170042P28 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: 30A, 250V PANEL-MOUNT FUSEHOLDER. FRU: Yes Status: Orderable Comments: FLS - FUSE HOLDER PART OF FMI KIT ** FE Assist ** IF FUSE HOLDER P/N 46-170042P28 FAILED, PLEASE ORDER MDAS FUSE FIX ON FAILURE KIT 5191271 (12/12/06 212001261)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.