Pular para o conteúdo principal
GE Healthcare · MRExcite22446072244605224486322445932244594

Will Not Scan with Gradient Errors.

Symptom

Will Not Scan with Gradient Errors.

Error message

2244607 The GP reported mutiple SGA Power Supply UnderVoltage Faults. same as 2244605 11X, G3, and OpenMR4. 2244607 The GP reported mutiple SGA Power Supply UnderVoltage Faults. 2244607 on 12X, 14X(HDx, HDe), E2, OpenMR5 sw is the same as 2244605 11X, G3, and OpenMR4. 2244607 The GP reported mutiple SGA Power Supply UnderVoltage Faults.

2244863 The %c Axis gradient amplifier failed to go to Standby when commanded. the same error as 2244593 on 11X, G3 and 2244594 on 12X, E2, 14M4 (HDx, HDe), and openMR4,5. 2244863 The %c Axis gradient amplifier failed to go to Standby when commanded. 2244863 error code on Post 14M5 is the same error as 2244593 on 11X, G3 and 2244594 on 12X, E2, 14M4 (HDx, HDe), and openMR4,5. 2244863 error code on Post 14M5 is the same error as 2244593 on 11X, G3 and 2244594 on 12X, E2, 14M4 (HDx, HDe), and openMR4,5.2244863 "The %c Axis gradient amplifier failed to go to Standby when commanded.->"

Applies to

Models
Excite
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 M5 Excite HDx HDi Platforms only.
Component
2334862-2 ACGD SGA PS ASM W/FRU PACKAGING See Long Description 2334862-2 Part Type: GP REPAIRABLES Part Weight: 2.18 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ACGD SGA PS ASM W/FRU PACKAGING FRU: Yes Status: Orderable This supply is for use with HFD, HFD-S, and ACGD Cabinets. ** WARNING - POSSIBLE PERSONAL INJURY. HEAVY OBJECT - THE POWER SUPPLY WITH CASE WEIGHS APPROXIMATELY 240 LBS (109 KG). Assistance to lift this product mechanical (Dollies) or Additional personnel] is required. 212002950 May 27, 2009
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues