GE Healthcare · CTLightSpeedVCT26013471490966426013471850015990773260134825
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Table will not move down; with tape or touch switch error See Long Description for switch locations. Table will not move down; tape switch was either contacted or not present The VCT table has a large number of touch sensors that can produce a variety of error codes.
Error message
Error: 260134714 Elevation touch switch2 (switches on underside of top side cover) is contacted. Open Long Description to see full error Thu Mar 17 20:37:24 2005 Host : Table Sub-System Ermes # : 260134714 Exception Class : Secondary Severity : Pri/Soft File : gtcb_emc_hardware.cxx Line# : 707 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Elevation touch switch2 (switches on underside of top side cover) is contacted. Position is 909664 um from the table base. Error: 260134718 Elevation touch switch2 (switches on underside of top side cover) isn't present. See Long Description Wed Oct 26 08:25:40 2005 Error: 260134718 Host: tgp Process: tgp_subsys File: gtcb_emc_hardware.cxx Line: 792 Wed Oct 26 08:25:40 2005 Host : Table Sub-System Ermes # : 260134718 Exception Class : Secondary Severity : Pri/Soft File : gtcb_emc_hardware.cxx Line# : 792 Function : No System Function Reported Scan Type : None/Unknown Scan: 50015/6/0 Elevation touch switch2 (switches on underside of top side cover) isn't present. Position is 990773 um from the table base. Error: 260134825 IMS cannot move because any of touch switches are already contacted or disconnected. See Long De
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5245271 GTCB3 (Global Table Control Board 3) This board replaces 5142177-3 GTCB2 and GTCB part # 5112470 thru 5112470-5, and 5142177. See long description for details Part #: 5245271 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: GTCB3 BOARD ASSY FRU: Yes Status: Orderable Comments: added per to ECO2080741 by sso#305010201 in May 8, 2009. Part will be available end of Sep, 2009, please order old p/n before new p/n on shelf.