GE Healthcare · CTHiSpeed18394965001
Scanner Hardware Stopped Scan
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 183949 Could NOT CLOSE Gantry (Axial) Loop Contactor See LD for full error CT01 Host: STC Proc: Axial Error: 183949 Function: AXIAL CONTROL : Axial Loop Contactor Interlocks Gantry (Axial) LOOP CONTACTOR Error. Could NOT CLOSE Gantry (Axial) Loop Contactor Another example: Tue May 24 08:41:21 2005 Suite: CT02 Host: STC Proc Axial Error: 183949 File: ax_error.c Method No Method 1114 Function: AXIAL CONTROL : Axial Loop Contactor Interlocks Scan: 65001/11/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 08:41:21:940 Log Series: 60 Gantry (Axial) LOOP CONTACTOR Error. Could NOT CLOSE Gantry (Axial) Loop Contactor. Axial Gantry Command/Status:0x9e28. EN 401 SR 442
Applies to
- Models
- HiSpeed
- Product
- HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-264806G1 STC AXIAL CONTROL BOARD See LD for details Part #: 46-264806G1 Part Type: GP REPAIRABLES Part Weight: 1.81 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: AXIAL CONTROL BOARD FRU: Yes Status: Orderable Comments: FLS - FE: READ SN T-1345. IF U455 ONAX BOARD IS BEING SHORTED, THECABLE MAY BE PINCHED. MS8365 4/95 HQ INV ONLY: PCN'S 161734 AND 161831 IMPLEMENTED. STOCK WASPURGED. 8/93 MBELL FLS - FE ASSIST: MS6588 MAY 2002 46-264806G1 = AXIAL BD FOR CT1.0 SEC. AXIAL ROTATION. THIS BOARD HAS THE PALS FOR 1.0 SECOND ROTATION. THE T3202BD =1.0 SEC AXIAL BD. THE T3202KD = .8 SEC AXIAL BD. THE ONLY DIFFERENCE BETWEEN THE T3202KD AND T3202BD ARE TWO PALS. THE T3202BD CAN BE UPGRADED TO THE T3202KD BY CHANGING PALS U211 AND U101. DO YOU STILL WANT TO ORDER THIS PART?? 46-264806G1 STC AXIAL CONTROL BOARD
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.