GE Healthcare · CTHiSpeed186403186507
Replace 2258540 collimator z-axis optical sensor
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 186403 OGP did not receive READY status from CIF for z-axis collimator control or OGP detected Aperture is not ready See LD for complete error Mon May 9 18:03:28 2005 Error: 186403 Host: CTE Process: tgp_out File: tgp_errors.c Line: 317 Aperture Error. Code : 20-1020-04 Error detected by : OGP OGP did not receive READY status from CIF for z-axis collimator control or OGP detected Aperture is not ready. Error: 186507 Z-Axis Collimator Error. Code : 30-1111-07 The number of pulses for Z-Axis Motor is under specification when moving from FWD limit to BWD limit on Home Positioning. See LD for complete error Mon May 9 18:03:19 2005 Error: 186507 Host: CTE Process: tgp_out File: tgp_in_input.c Line: 449 Z-Axis Collimator Error. Code : 30-1111-07 Error detected by : CIF The number of pulses for Z-Axis Motor is under specification when moving from FWD limit to BWD limit on Home Positioning.
Applies to
- Models
- HiSpeed
- Product
- CT/e Dual, CT ProSpeed AII/FII Irix or Linux HiSpeed CT/e Dual CT Scanner System see long Description for sales information Product Name: CT/e Dual CT Scanner System Irix A list of types. (a Product/hardware table) N/A An On-site / Remote identification aid . (Example: check_config) N/A A Table of options (included, optional, new introductions, …) N/A A list of Software that has run on this product (Product / Software Compatibility Table) N/A Product photographs N/A Marketing Information N/A Documents: Always use the most appropriate documentation for your system. Documentation can be found on CT Service CDROM/DVD, or online at FCT Service Engineering Tech Pubs Website, Common Document Library, and My Workshop. Links: Drawing Print 2320306-2CCT : http://gemspdm.med.ge.com/ematrix/common/eLibDocument.jsp?type=type_DrawingPrint&name=2320306-2CCT&rev=9 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2258540 P2014NJ (also 2258540-2) OPTO ASSY - Used on YMS HiSpeed Collimator See long description for details Part #: 2258540 Part Type: GP CONSUMABLES Part Weight: 0.04 KGM Repairable: No Returnable: No Hazardous Material: N Description: P2014NJ SNSR ASSY FRU: Yes Status: Orderable ====================== Parts Offered as: Current: 2258540 Harvested: Not Available Exchange: Not Available Comments: ** FE ASSIST ** For CT Hispeed Dual For GDAS Gantry, 2258540-2 can replace 2258540. 2258540-2 is backward compatible with 2258540. For MDAS Gantry, 2258540-2 and 5120153-2 should be used together to replace 2258540 (212001261 7/13/08) FLS - Z AXIS SENSOR ASM U/O EBISU (NXI) GANRY MS08992 6/5/00
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required