Unable to Scan Using Invivo 7-Channel Breast Coil Due to Wrong 2_BREAST Coil Config.
Symptom
Unable to Scan Using Invivo 7-Channel Breast Coil Customer unable to scan using the Invivo 7-Channel Breast Coil. When they try this they receive the error message shown in the Long Description. Note that Invivo's Operator manual for this coil used to direct the customer to use a modified "2_BREAST" coil config for this coil. A newer revision of their manual has different instructions. Also please note that GE does not service this coil as it is sold directly by Invivo to the customer. Customer unable to scan using the Invivo 7-Channel Breast Coil. When they try this they receive the error message shown below. Note that Invivo's operator manual for this coil used to direct the customer to use a modified "2_BREAST" coil config for this coil. A newer revision of their manual has different instructions. Also please note that GE does not service this coil as it is sold directly by Invivo to the customer. Error: 2220007 Host: mr02 Process: scn File: RxMgrDownload.m Line: 2143 The prescribed RF coil and detected RF coil do not match. Upper coil connect status:0, Upper coil code:Null, Lower coil connect status:1, Lower coil cod e:WFdtKBdrLUmJKY398qJocJG3XcRNtyYd, Rx coil name:2_BREAST, Rx coil code:p1vf g3d42e3afd74fnok9d3cc310526v
Error message
2220007 The prescribed RF coil and detected RF coil do not match. See long description on reading/interpreting this type of error message Note: An example of this error message, with explanation, is provided below. Wed Mar 3 08:18:58 2004 Error: 2220007 Host: lx-mr3 Process: scn File: RxMgrDownload.m Line: 2074 The prescribed RF coil and detected RF coil do not match. 040303104912225 Upper coil connect status:0, Upper coil code:Null, Lower coil connect status:1, Lower coil code:q8dCUW0PD2EaETIplAp90PLrACM6CW38, Rx coil name:MRI 8HRKNRR PA, Rx coil code:q8dCUWOPD2EaETIpIAp9oPLrACM6CW38 NOTE: Upper coil connector status refers to connector plugged into upper port at "dog house" (i.e. head coil or Med' Adv' QUAD KNEE ("chimney") coil). Lower coil connect status refers to all PA coils / any coil using the lower port at the "dog house". Coil code should match that of what was detected (in this case the zeros were config'd as O and o (oh's - upper and lower case), and the number one following "ETIp" was an "I").Applies to
- Models
- Excite, Signa
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.