Scan aborts with errors 183180, 183337, 183338

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 183180 I/O board fault detected. Relay K245 open, HVDC Bus not active. Cannot close backup contactor. EEBCSTAT = 1 KILLBC = 0 See Long Description for complete error Sun Jun 5 23:17:07 2005 Error: 183180 Host: OBC Process: Generator File: gen_rotor.c 1.39 Line: 450 Function: SYSTEM POWER CONTROL : HVDC Control Scan: 33003/1/1 Type: Scout Exception Level: Pri/Most Severe Time: 23:17:06:735 Log Series: 425 I/O board fault detected. Relay K245 open, HVDC Bus not active. Cannot close backup contactor. EEBCSTAT = 1 KILLBC = 0.

Error: 183337 The HVDC Bus voltage has dropped below the operating level See Long Description for Details. Tue Sep 4 12:15:09 2001 Error: 183337 Suite: CT Host: OBC Process: Generator File: gen_power.c 1.13 Method: No Method Line: 207 Function: SYSTEM MONITORING : 550V Monitor Scan: 1/1/1 Type: Scout Exception Level: Pri/Most Severe Time: 12:15:08:835 Log Series: 122 The HVDC Bus voltage has dropped below the operating level. No load HVDC value: 658V HVDC rail value: 464V Rotor LOV bit In the errorlog.

Error: 183338 The maximum allowable HVDC Rail voltage drop has been exceeded. See LD for complete error Tue Jul 25 09:43:35 2006 Error: 183338 Host: OBC Pr

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 1.X - GRE H1_P_S4 Lightspeed 1.X - GRE Method of visual identification Large square gantry similar in appearance to a CTi gantry. Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2168593 HEMRC I/F FILTER BOARD See long descriptions for details. Part #: 2168593 Part Type: GP CONSUMABLES Part Weight: 1.81 LBR Repairable: No Returnable: No Hazardous Material: N Description: HEMRC IF FILTER FRU: Yes Status: Orderable Comments: FLS - HEMRC I/F FILTER USED ON FAST RHAPSODE 6/27/97 M3 FW28 GWENN
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.