Pular para o conteúdo principal
GE Healthcare · CTLightSpeed230017821470544300000230017801

Collimator calibration drift check has failed Cal will continue with the next step - FPA Check.

Symptom

Collimator calibration drift check has failed Cal will continue with the next step - FPA Check.

Error message

Error: 230017821 ACCS CCal: Drift check: The polynomial value has drifted by 0.470544 See the LD for the complete error code Error: 230017821 ACCS CCal: Drift check: The polynomial value has drifted Wed Aug 9 08:18:38 2006 Error: 230017821 Host: qxi1_sbc Process: accs File: collSweepProcessing.c Line: 1635 ACCS CCal: Drift check: The polynomial value has drifted by 0.470544 which is more than tolerance 0.300000

Error: 230017801 ACCS CCal: Sweep Processing failed. Status = -1 See the Long Description for the full error code Tue Nov 30 02:33:04 2004 Error: 230017801 Host: k2_ct2_ls Process: accs File: ccal_prep.c Line: 661 ACCS CCal: Sweep Processing failed. Status = -1

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues