Pular para o conteúdo principal
GE Healthcare · CTLightSpeed260118315

LightSpeed 5.X (60kW)

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260118315 JEDI ID Communication Timer Expired. Check power and commnication interconnects. See Long Description for complete error Mon Dec 6 12:04:49 2004 Error: 260118315 Host: orp Process: orp_subsys File: gen_exec.c Line: 1665 Mon Dec 6 12:05:01 2004 Host : On-Board Rotating Processor Ermes # : 260118315 Exception Class : Soft Severity : Pri/Soft File : gen_exec.c Line# : 1665 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 JEDI ID Communication Timer Expired. Check power and commnication interconnects.

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2409042 (new pn) 2325360-6 (old pn) LVPS 400 (250V) used in JEDI Aux Box of all Xi HiSpeeds and 5.X RT, 60kW 100kW LightSpeeds, Brightspeed 4/8/16 See LD for FEMC info and picture 2409042 (new pn) 2325360-6 (old pn) LVPS 400 (250V) used in JEDI Aux Box of all Xi HiSpeeds and 5.X RT, 60kW 100kW LightSpeeds picture below Note: This Item is listed as a FRU YES for the HP60 but even though the part number is the same, it is listed as a FRU NO for HP100 and RT. Part #: 2325360-6 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: LVPS 400 (250V) BOOT 2.05/APPLI 2.36 FRU: Yes Status: Orderable Comments: FLS - LVPS 400 (250V) BOOT USED ON: GENERATOR, JEDI 100 2.05/APPLI 2.36 VASC 1T ADDED PER SPCN: SUPPLIED BY GEMS INDIA 030903RF141542N1 MS13452 030908 RISK: SITE DOWN FREQUENCY: MEDIUM PER SPCN THIS PART SERVES AS REPLACEMENT FOR 2325360-5, ALL REMAINING STOCK OF -5 IS TO BE UPGRADED TO -6 VERSION. The only difference between pn 2409042 (new pn) and pn 2325360-6 (old pn) is packaging improvment . You can order pn 2325360-6 if no pole stock on pn 2409042 (6/2005 P.Liebau)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required